Test space capacity in litres:
- 5 K/min 200 / 350 / 600 / 1000 / 1500
- 10 K/min and 15 K/min 200 / 350 / 500 / 810 / 900 / 1250
Temperature range:
-40 °C / +180 °C -70 °C / +180 °C
Temperature change rates:
5 to 30 K/min
Design:
Control: microprocessor control and monitoring system Test space: stainless steel grade 1.4301 Refrigerants: chloride free, hermetically sealed Humidity system (only type CS): capacitive measuring
Options:
- Additional entry ports ø 50 mm, 80 mm und 125 mm
- Additional shelves
- Panoramic window, multiple glazed, sized 570 x 710 mm
- Glove box hand-hole in the door
- Additional Pt 100 for temperature measuring on test specimen alternative reversible to controller function
- Air dryer system
- Mobile design – wheels
- Temperature protection for test-specimen
- Water cooled compressor; external condenser
- CID software for programming and documentation
- Variable air speed in the test space
- RS 232 interface changeable to USB
- ETHERNET interface RJ 45
Other chamber sizes and options on request.
Type
|
5 K/min
|
|
TS -40/…, CS -40/… TS -70/…, CS -70/…
|
-40/200-5 -70/200-5
|
-40/350-5 -70/350-5
|
-40/600-5 -70/600-5
|
-40/1000-5 -70/1000-5
|
-40/1500-5 -70/1500-5
|
Test Space |
liters
|
200
|
350
|
600
|
1000
|
1500
|
Test Space Dimensions |
|
height |
mm, appr.
|
750
|
750
|
850
|
900
|
900
|
|
width |
mm, appr.
|
650
|
650
|
850
|
1000
|
1000
|
|
depth |
mm, appr.
|
400
|
720
|
830
|
1100
|
1600
|
Overall Dimensions |
|
height |
mm, appr.
|
1800
|
1800
|
1900
|
1950
|
1950
|
|
width |
mm, appr.
|
920
|
920
|
1120
|
1270
|
1270
|
|
depth |
mm, appr.
|
1345
|
1665
|
1875
|
2245
|
2745
|
Temperature Tests
|
Temperature Range |
TS -40/…, CS -40/… |
°C
|
-40/+180
|
-40/+180
|
-40/+180
|
-40/+180
|
-40/+180
|
TS -70/…, CS -70/… |
°C
|
-70/+180
|
-70/+180
|
-70/+180
|
-70/+180
|
-70/+180
|
Temp. Change Rate Cooling |
TS -40/…, CS -40/… |
K/min
|
5,0
|
5,0
|
7,0
|
5,0
|
5,0
|
acc. IEC 60068-3-5 |
TS -70/…, CS -70/… |
K/min
|
5,0
|
5,0
|
7,0
|
5,0
|
5,0
|
Temp. Change Rate Heating |
TS -40/…, CS -40/… |
K/min
|
8,0
|
8,0
|
6,0
|
6,0
|
5,0
|
acc. IEC 60068-3-5 |
TS -70/…, CS -70/… |
K/min
|
8,0
|
8,0
|
6,0
|
6,0
|
5,0
|
max. Thermal Load at +20°C |
TS -40/…, CS -40/… |
W
|
3500
|
3500
|
4500
|
5000
|
6000
|
TS -70/…, CS -70/… |
W
|
3500
|
3500
|
5000
|
5500
|
5000
|
Temperature Fluctuation |
K
|
≤± 0,3 temporally
|
Climatic Tests
|
Temperature Range |
CS -40/…, CS -70/… |
°C
|
+10 to +95
|
Humidity Range |
CS -40/…, CS -70/… |
% RH
|
10 to 98
|
Dew Point Range I |
CS -40/…, CS -70/… |
°C
|
+7 to +94
|
Dew Point Range II |
CS -40/…, CS -70/… |
°C
|
-10 to +7
|
Humidity Fluctuation I |
CS -40/…, CS -70/… |
% RH
|
≤± 1,5
|
Humidity Fluctuation II |
CS -40/…, CS -70/… |
% RH
|
≤± 3
|
Temperature Fluctuation |
CS -40/…, CS -70/… |
K
|
≤± 0,3
|
Nominal Voltage |
400 V +6% -10%, 3/N, 50 Hz
|
Nominal Power |
TS -40/… |
kW, appr.
|
9,3
|
9,3
|
9,2
|
13,3
|
13,2
|
CS -40/… |
kW, appr.
|
10,4
|
10,4
|
12,1
|
13,2
|
13,2
|
TS -70/… |
kW, appr.
|
10,9
|
10,9
|
14,2
|
16,4
|
16,4
|
CS -70/… |
kW, appr.
|
12,0
|
12,0
|
14,2
|
16,4
|
16,4
|
Condenser |
|
air-cooled
|
water-cooled
|
Weight |
TS -40/…, CS -40/… |
kg, appr.
|
500
|
600
|
700
|
850
|
1000
|
TS -40/…, CS -40/… |
kg, appr.
|
550
|
650
|
750
|
900
|
1100
|
Noise level |
|
dB(A)
|
< 58
|
< 58
|
< 62
|
< 62
|
< 62
|
Type
|
10 K/min
|
|
TS -40/…, CS -40/… TS -70/…, CS -70/…
|
-40/200-10 -70/200-10
|
-40/350-10 -70/350-10
|
-40/500-10 -70/500-10
|
-40/900-10 -70/900-10
|
-40/1250-10 -70/1250-10
|
Test Space |
liters
|
200
|
350
|
500
|
900
|
1250
|
Test Space Dimensions |
|
height |
mm, appr.
|
750
|
750
|
820
|
870
|
870
|
|
width |
mm, appr.
|
650
|
650
|
850
|
1000
|
1000
|
|
depth |
mm, appr.
|
400
|
720
|
725
|
1000
|
1400
|
Overall Dimensions |
|
height |
mm, appr.
|
1800
|
1800
|
2170
|
2230
|
2230
|
|
width |
mm, appr.
|
920
|
920
|
1120
|
1270
|
1270
|
|
depth |
mm, appr.
|
1345
|
1665
|
2375
|
2645
|
3145
|
Temperature Tests
|
Temperature Range |
TS -40/…, CS -40/… |
°C
|
-40/+180
|
-40/+180
|
-40/+180
|
-40/+180
|
-40/+180
|
TS -70/…, CS -70/… |
°C
|
-70/+180
|
-70/+180
|
-70/+180
|
-70/+180
|
-70/+180
|
Temp. Change Rate Cooling |
TS -40/…, CS -40/… |
K/min
|
10,0
|
10,0
|
10,0
|
10,0
|
10,0
|
acc. IEC 60068-3-5 |
TS -70/…, CS -70/… |
K/min
|
11,0
|
10,0
|
10,0
|
10,0
|
10,0
|
Temp. Change Rate Heating |
TS -40/…, CS -40/… |
K/min
|
12,0
|
10,0
|
10,0
|
10,0
|
10,0
|
acc. IEC 60068-3-5 |
TS -70/…, CS -70/… |
K/min
|
12,0
|
10,0
|
10,0
|
10,0
|
10,0
|
max. Thermal Load at +20°C |
TS -40/…, CS -40/… |
W
|
5000
|
5000
|
7000
|
8000
|
8000
|
TS -70/…, CS -70/… |
W
|
4000
|
5000
|
5500
|
8000
|
8000
|
Temperature Fluctuation |
K
|
≤± 0,3 temporally
|
Climatic Tests
|
Temperature Range |
CS -40/…, CS -70/… |
°C
|
+10 to +95
|
Humidity Range |
CS -40/…, CS -70/… |
% RH
|
10 to 98
|
Dew Point Range I |
CS -40/…, CS -70/… |
°C
|
+7 to +94
|
Dew Point Range II |
CS -40/…, CS -70/… |
°C
|
-10 to +7
|
Humidity Fluctuation I |
CS -40/…, CS -70/… |
% RH
|
≤± 1,5
|
Humidity Fluctuation II |
CS -40/…, CS -70/… |
% RH
|
≤± 3
|
Temperature Fluctuation |
CS -40/…, CS -70/… |
K
|
≤± 0,3
|
Nominal Voltage |
400 V +6% -10%, 3/N, 50 Hz
|
Nominal Power |
TS -40/… |
kW, appr.
|
13,4
|
13,4
|
17,5
|
23,0
|
32,0
|
CS -40/… |
kW, appr.
|
13,4
|
13,4
|
18,6
|
24,1
|
33,0
|
TS -70/… |
kW, appr.
|
16,6
|
16,6
|
20,3
|
26,0
|
37,0
|
CS -70/… |
kW, appr.
|
16,6
|
16,6
|
20,3
|
26,0
|
35,8
|
Condenser |
|
water-cooled
|
Weight |
TS -40/…, CS -40/… |
kg, appr.
|
600
|
700
|
800
|
1000
|
1150
|
TS -40/…, CS -40/… |
kg, appr.
|
650
|
750
|
900
|
1100
|
1300
|
Noise level |
|
dB(A)
|
< 58
|
< 58
|
< 68
|
< 68
|
< 73
|
Type
|
15 K/min
|
|
TS -40/…, CS -40/… TS -70/…, CS -70/…
|
-40/280-15 -70/280-15
|
-40/500-15 -70/500-15
|
-40/810-15 -70/810-15
|
-40/1250-5 -70/1250-15
|
Test Space |
liters
|
280
|
500
|
810
|
1250
|
Test Space Dimensions |
|
height |
mm, appr.
|
720
|
820
|
870
|
870
|
|
width |
mm, appr.
|
650
|
850
|
1000
|
1000
|
|
depth |
mm, appr.
|
615
|
730
|
900
|
1400
|
Overall Dimensions |
|
height |
mm, appr.
|
2080
|
2170
|
2230
|
2230
|
|
width |
mm, appr.
|
920
|
1120
|
1270
|
1270
|
|
depth |
mm, appr.
|
2165
|
2375
|
2645
|
3145
|
Temperature Tests
|
Temperature Range |
TS -40/…, CS -40/… |
°C
|
-40/+180
|
-40/+180
|
-40/+180
|
-40/+180
|
TS -70/…, CS -70/… |
°C
|
-70/+180
|
-70/+180
|
-70/+180
|
-70/+180
|
Temp. Change Rate Cooling |
TS -40/…, CS -40/… |
K/min
|
15,0
|
15,0
|
15,0
|
15,0
|
acc. IEC 60068-3-5 |
TS -70/…, CS -70/… |
K/min
|
15,0
|
15,0
|
15,0
|
15,0
|
Temp. Change Rate Heating |
TS -40/…, CS -40/… |
K/min
|
15,0
|
15,0
|
15,0
|
15,0
|
acc. IEC 60068-3-5 |
TS -70/…, CS -70/… |
K/min
|
15,0
|
15,0
|
15,0
|
15,0
|
max. Thermal Load at +20°C |
TS -40/…, CS -40/… |
W
|
5000
|
8000
|
8000
|
8000
|
TS -70/…, CS -70/… |
W
|
5000
|
8000
|
8000
|
8000
|
Temperature Fluctuation |
K
|
≤± 0,3 temporally
|
Climatic Tests
|
Temperature Range |
CS -40/…, CS -70/… |
°C
|
+10 to +95
|
Humidity Range |
CS -40/…, CS -70/… |
% RH
|
10 to 98
|
Dew Point Range I |
CS -40/…, CS -70/… |
°C
|
+7 to +94
|
Dew Point Range II |
CS -40/…, CS -70/… |
°C
|
-10 to +7
|
Humidity Fluctuation I |
CS -40/…, CS -70/… |
% RH
|
≤± 1,5
|
Humidity Fluctuation II |
CS -40/…, CS -70/… |
% RH
|
≤± 3
|
Temperature Fluctuation |
CS -40/…, CS -70/… |
K
|
≤± 0,3
|
Nominal Voltage |
400 V +6% -10%, 3/N, 50 Hz
|
Nominal Power |
TS -40/… |
kW, appr.
|
17,6
|
24,0
|
32,0
|
42,5
|
CS -40/… |
kW, appr.
|
18,7
|
24,0
|
33,0
|
43,6
|
TS -70/… |
kW, appr.
|
21,2
|
27,4
|
37,1
|
46,8
|
CS -70/… |
kW, appr.
|
21,2
|
27,4
|
35,8
|
46,8
|
Condenser |
|
water-cooled
|
Weight |
TS -40/…, CS -40/… |
kg, appr.
|
800
|
900
|
1100
|
1350
|
TS -40/…, CS -40/… |
kg, appr.
|
900
|
1000
|
1200
|
1500
|
Noise level |
|
dB(A)
|
< 68
|
< 68
|
< 73
|
< 75
|
Temperature Change Rate acc. IEC 60068-3-5:
|
heating |
= |
Δ th / Δ Th |
cooling |
= |
Δ tk / Δ Tk |
Δ Th = time of heating Δ Tk = time of cooling Δ th = temperature difference, heating Δ tk = temperature difference, cooling All figures are average values which have been obtained at a temperature of +25° C, without test specimens, without thermal load and without options. We reserve the right to make alterations due to technical development.